| Model Number |
N/A |
| Medical Device Problem Code |
Separation Problem (4043)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).D10: blunt tip screw, ã¿ 4x46mm item # 47248604640 lot# 3185193, ann blunt tip screw 4x42mm item # 47248604240 lot# 3189213, ann blunt tip screw 4x44mm item # 47248604440 lot# 3193251, ann blunt tip screw 4x56mm item # 47248605640 lot# 3181925, ann cort bone screw 4 x 24mm item # 47248612440 lot# 3185200, ann cort bone screw 4 x 30mm item # 47248613040 lot# 3177549.G2: foreign: japan.The device will not be returned for analysis; however, an investigation of the reported event is in progress.Once the investigation is completed, a supplemental mdr will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that the surgeon found that the #1 proximal screw had migrated out of the proper position, approximately 5 weeks after initial implantation.Revision surgery is being considered to remove the migrated screw, but no specific date has been set.There is no reported harm or injury to the patient, however this malfunction has caused a serious injury in the past.Due diligence has been completed for this complaint; to date all available additional information has been received.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay additional information.The following sections were updated: b4, g3, g6, h2, h3, h6, h10, h11.No product was returned or pictures provided; visual and dimensional evaluations could not be performed.A review of the device manufacturing records confirmed no abnormalities or deviations.Review of the complaint histories identified additional similar complaints for the reported items and no additional similar complaints for the reported part and lot combinations.Complaints are monitored per complaint trending process in order to identify potential adverse trends.It can be assumed that multiple contributing factors, related to either patient condition and behavior or implantation procedure, contributed to the migration of the screws.If and to what extent any of these aspects may have influenced the migration of the screw remains unknown.An additional deeper investigation was performed which identified the design limitation of the corelock mechanism of the affixus nail as a potential contributing factor.As part of the deeper investigation, a scientific literature search was conducted and a systematic review of the outcome of intramedullary nailing for acute proximal humerus fracture showed that screw migration and perforation into the joint is the second most common complication following secondary loss of reduction.In addition, scientific literature of competitor and predicate devices were assessed.This review has shown that the affixus® natural nail® proximal humeral system performs better and/or in equal range in comparison with legally marketed similar devices.In conclusion, as the cause may be multifactorial an exact root cause could not be identified for the migration of the screw.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|
Event or Problem Description
|
|
No additional event information to report at this time.
|
| |
|