ARTHREX, INC. REDEUCE PUMP TUBING W/CONN, 8' LONG; ARTHROSCOPE AND ACCESSORIES Back to Search Results
Model Number REDEUCE PUMP TUBING W/CONN, 8' LONG
Medical Device Problem Code Excess Flow or Over-Infusion (1311)
Health Effect - Clinical Code Extravasation (1842)
Date of Event 08/09/2024
Type of Reportable Event Malfunction
Event or Problem Description
On 08/09/2024, it was reported by a sales representative via (b)(4) that an ar-6480 dualwave pump was acting up and pushing a ton of fluid into the patient's joint space.This occurred during a knee arthroscopy.The pump finally evened out and worked as usual.The fluid was more than usual in the knee post-op.Additional information was provided on 08/13/2024, where it was stated by the sales representative that this event occurred on 08/07/2024, and an ar-6410 tubing was used.The tubing was replaced, and there were no further issues.The patient did have fluid in the knee, and they just put dressings on and the knee fluid came down.The patient was discharged that day.
 
Additional Manufacturer Narrative
Investigation is in process.A follow-up report will be provided upon availability of additional information.
 
Additional Manufacturer Narrative
Additional information: g3, h3, h6.Based on the information provided which may include the device (if available and returned), pictures, videos, event description, and any additional information from the field, arthrex was able to conclude a most likely cause.The most likely cause for the reported failure can be attributed to user error due to unplugging and plugging the pressure line connector back into the arthroscopy pump, thereby creating a pressure decay on the pump and/or spiking the bags of fluid and allowing that fluid to migrate through the tubing before plugging the pressure line connector into the pump.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
REDEUCE PUMP TUBING W/CONN, 8' LONG
Common Device Name
ARTHROSCOPE AND ACCESSORIES
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer Contact
jared engle
8009337001
MDR Report Key20124416
Report Number1220246-2024-07332
Device Sequence Number3152061
Product Code HRX
UDI-Device Identifier00888867118027
UDI-Public00888867118027
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K083707
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial,Followup
Report Date (Section B) 07/03/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberREDEUCE PUMP TUBING W/CONN, 8' LONG
Device Catalogue NumberAR-6411
Was Device Available for Evaluation? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/09/2024
Supplement Date Received by Manufacturer06/10/2025
Initial Report FDA Received Date09/03/2024
Supplement Report FDA Received Date07/03/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1