• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: W. L. GORE & ASSOCIATES, INC. GORE® TAG® CONFORMABLE THORACIC STENT GRAFT WITH ACTIVE CONTROL SYSTEM; SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

W. L. GORE & ASSOCIATES, INC. GORE® TAG® CONFORMABLE THORACIC STENT GRAFT WITH ACTIVE CONTROL SYSTEM; SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT Back to Search Results
Catalog Number TGM454515
Medical Device Problem Code Patient-Device Incompatibility (2682)
Health Effect - Clinical Code Clinical Endoleak (4843)
Date of Event 08/08/2024
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
W.L.Gore & associates, inc.(gore) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by gore, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Blank fields present on this report include required fields and fields determined to be not applicable.Blank required fields indicate that the information was not provided, was deemed unavailable or was not applicable.This report does not constitute an admission or a conclusion by fda, gore, or its associates that the device, gore or its associates caused or contributed to the event described in the report.In particular, this report does not constitute a legal admission by anyone that the product described in this report has any defects or has malfunctioned, as defined from a legal standpoint.These words are included in the report and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Event or Problem Description
The following was reported to gore: in (b)(6) 2022, a gore® tag® conformable thoracic stent graft (ctag) was implanted.On (b)(6) 2024, a gore associate was notified about a patient who had a type ia endoleak in the ascending aorta.On (b)(6) 2024, a reintervention was performed, extending with a ctag device.
 
Additional Manufacturer Narrative
H.6.Investigation findings code c20: the device remains implanted and is not available for analysis.H.6.Investigation findings code c19: a review of the manufacturing records for the device verified that the lot met all pre-release specifications.H.6.Investigation conclusions code d12: it should be noted that, per the gore® tag® conformable thoracic stent grafts with active control system instructions for use, complications associated with the use of the gore® tag® conformable thoracic stent graft may include, but are not limited to, endoleak.Updated h.6.Type of investigation from b21 to b13, b14, b20.Updated h.6.Investigation findings from c21 to c19 and c20.Updated h.6.Investigation conclusions from d16 to d15 and d12.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GORE® TAG® CONFORMABLE THORACIC STENT GRAFT WITH ACTIVE CONTROL SYSTEM
Common Device Name
SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT
Manufacturer (Section D)
W. L. GORE & ASSOCIATES, INC.
1505 n. fourth street
flagstaff AZ 86004
Manufacturer (Section G)
MEDICAL WOODY SPRINGS B/P
3450 w. kiltie lane
flagstaff AZ 86005
Manufacturer Contact
bryce wargin
1505 n. fourth street
flagstaff, AZ 86004
9285263030
MDR Report Key20125912
Report Number2017233-2024-05279
Device Sequence Number3187253
Product Code MIH
UDI-Device Identifier00733132646999
UDI-Public00733132646999
Combination Product (Y/N)N
PMA/510(K) Number
P040043
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/24/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date02/12/2023
Device Catalogue NumberTGM454515
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/08/2024
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date09/03/2024
Supplement Report FDA Received Date10/24/2024
Date Device Manufactured02/13/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age70 YR
Patient SexMale
-
-