Model Number N/A
Medical Device Problem Code Separation Failure (2547)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/23/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).D10: 110003453; item name: g7 curved acet shell inserter; lot # 401800, unk ball hex driver.The product has been received by zimmer biomet.The investigation is currently in process.Once the investigation has been completed, a follow-up report will be submitted.
 
Event or Problem Description
It was reported that the inserter screw wouldn't release from the inserter during surgery.The doctor tried the ball hex driver to remove with such force that it broke off.The procedure was completed using a monoblock handle.There is no additional information available at the time of this report.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.Visual examination of the returned product/provided pictures identified impact marks visible to the strike plate of the inserter along with damage to the rubber handle and scuffing to the shaft.The inserter shaft has a fractured hex head stuck inside of the device.The rest of the driver was not returned, and the head could not be removed for evaluation.Review of the device history record identified no deviations or anomalies during manufacturing related to the reported event.Medical records were not provided.A definitive root cause cannot be determined.Based on product review, complaint is confirmed.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
There is no update to the prior event description provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
G7 CURVED INSERTER THD SHAFT
Common Device Name
INSTRUMENT, HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key20126262
Report Number0001825034-2024-02133
Device Sequence Number12616760
Product Code LXH
UDI-Device Identifier00887868462528
UDI-Public(01)00887868462528(11)171207(10)053479
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/22/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue Number110003454
Device Lot Number053479
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/05/2024
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/21/2024
Supplement Date Received by Manufacturer10/21/2024
Initial Report FDA Received Date09/03/2024
Supplement Report FDA Received Date10/24/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/07/2017
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexMale
-
-