Catalog Number 367925
Medical Device Problem Code Short Fill (1575)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/02/2024
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that while using bd vacutainer® sodium fluoride/potassium oxalate 15mg/12mg, an unspecified amount of tube underfilled.No patient impact reported.
 
Additional Manufacturer Narrative
E1: initial reporter phone #: (b)(6).A device evaluation and/or device history review is anticipated but is not complete.Upon completion, a supplemental report will be filed.
 
Event or Problem Description
It was reported that while using bd vacutainer® sodium fluoride/potassium oxalate 15mg/12mg, an unspecified amount of tube underfilled.No patient impact reported.
 
Additional Manufacturer Narrative
Investigation summary: bd had not received samples, but 1 photo was provided for investigation.The photo was reviewed and the indicated failure mode for underfill was observed.Additionally, 100 retention samples from bd inventory were evaluated by both visual examination and functional testing and the issue of underfill was not observed.Based on a review of the device history record for the incident lot, all product specifications and requirements for lot release were met.There were no related quality issues during manufacturing of the product.This complaint has been confirmed for the indicated failure mode of underfill based on photo analysis.Bd was not able to identify a root cause for the indicated failure mode complaints received for this device and reported condition will continue to be tracked and trended.Our business team regularly reviews the collected data for identification of emerging trends.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD VACUTAINER® SODIUM FLUORIDE/POTASSIUM OXALATE 15MG/12MG
Common Device Name
BLOOD SPECIMEN COLLECTION DEVICE
Manufacturer (Section D)
BECTON, DICKINSON & CO. (BROKEN BOW)
150 south 1st avenue
broken bow NE 68822
Manufacturer (Section G)
BECTON, DICKINSON & CO. (BROKEN BOW)
150 south 1st avenue
broken bow NE 68822
Manufacturer Contact
jo doyka
7 loveton circle
sparks, MD 21152
4103164000
MDR Report Key20127278
Report Number1917413-2024-00811
Device Sequence Number7417825
Product Code JKA
UDI-Device Identifier30382903679257
UDI-Public(01)30382903679257
Combination Product (Y/N)N
Initial Reporter CountryCO
PMA/510(K) Number
K231237
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 09/17/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date03/31/2025
Device Catalogue Number367925
Device Lot Number3290580
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/05/2024
Supplement Date Received by Manufacturer09/17/2024
Initial Report FDA Received Date09/03/2024
Supplement Report FDA Received Date09/25/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured10/17/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1