| Model Number |
V173 |
| Medical Device Problem Code |
Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/14/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that this cardiac resynchronization therapy pacemaker (crt-p) was found to be in safety mode.The device is schedule to be explanted.No adverse patient effects were reported.Additional information was received, mentioning that this device was successfully explanted and replaced due to safety switch.The device is expected to return for analysis.No additional adverse patient effects were reported.
|
| |
|
Event or Problem Description
|
|
It was reported that this cardiac resynchronization therapy pacemaker (crt-p) was found to be in safety mode.The device is schedule to be explanted.No adverse patient effects were reported.Additional information was received, mentioning that this device was successfully explanted and replaced due to safety switch.The device is expected to return for analysis.No additional adverse patient effects were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Upon receipt at our post market quality assurance laboratory, this device was thoroughly inspected and analyzed.Interrogation of the device confirmed it was operating in safety mode due to system resets.It was confirmed that brady therapy remained available.The high internal impedance resulted in the system resets due to temporary high-power consumption and subsequent reversion to safety mode operation.Boston scientific has issued a field safety notice regarding ingenio extended life (el) and cardiac resynchronization therapy pacemaker (crt-p) devices that may exhibit this behavior.This particular device is included in the high impedance in ingenio extended life (el) pacemaker and crt-p advisory population.
|
| |
|
Event or Problem Description
|
|
It was reported that this cardiac resynchronization therapy pacemaker (crt-p) was found to be in safety mode.The device is schedule to be explanted.No adverse patient effects were reported.Additional information was received, mentioning that this device was successfully explanted and replaced due to safety switch.The device is expected to return for analysis.No additional adverse patient effects were reported.
|
| |
|