| Model Number |
WA22355C |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/26/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The evaluation of the event is ongoing.Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|
Event or Problem Description
|
|
It was observed during the device inspection, that the high frequency resection electrode exhibited the distal end wire was broken.There were no reports of patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being supplemented to provide additional information based on the legal manufacturer's investigation.Based on the results of the investigation, it is likely that the following led to the malfunction: excessive external force or fracture due to frequency loss.A definitive root cause cannot be identified.A device history review revealed no issues that could have caused or contributed to the reported issue.Olympus will continue to monitor the field performance of this device.Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|