W. L. GORE & ASSOCIATES, INC. GORE® EXCLUDER® AAA ENDOPROSTHESIS; SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT
|
Back to Search Results |
|
| Catalog Number |
PLC271400 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
Clinical Endoleak (4843)
|
| Date of Event |
08/08/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The instructions for use (ifu) for the gore® excluder® aaa endoprosthesis states; adverse events that may occur and / or require intervention include but are not limited to: endoleak, w.L.Gore & associates, inc.(gore) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by gore, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Blank fields present on this report include required fields and fields determined to be not applicable.Blank required fields indicate that the information was not provided, was deemed unavailable or was not applicable.This report does not constitute an admission or a conclusion by fda, gore, or its associates that the device, gore or its associates caused or contributed to the event described in the report.In particular, this report does not constitute a legal admission by anyone that the product described in this report has any defects or has malfunctioned, as defined from a legal standpoint.These words are included in the report and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.This statement should be included with any information or report disclosed to the public under the freedom of information act.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2024, this patient underwent endovascular treatment for a zone 5 descending aortic aneurysm.The patient was implanted with gore® excluder® abdominal branch endoprosthesis (tambe).The complete tambe system is comprised of the tambe aortic component, the branch components (gore® viabahn® (vbx) balloon expandable endoprosthesis), a distal bifurcated component (gore® excluder® iliac branch endoprosthesis component) and contralateral legs (gore® excluder® contralateral legs).At the conclusion of the procedure a slight distal type i endoleak was noted and determined to be associated with the gore® excluder® contralteral leg component (plc271400/28148793) in the left common iliac artery.The physician chose to watch and wait to see if the endoleak would resolve itself.The patient tolerated the procedure.On an unknown date the patient underwent follow-up imaging and the patient was determined to have a pseudoaneurym in the brachial artery, which is believed to be procedure, not device related.On (b)(6) 2024, the patient underwent re-intervention and the gore® excluder® contralteral leg components in the common iliac artery extended from the tambe system were bilaterally extended with (please note: distal type i endoleak only on left side but both sides were extended) viabahn® (vbx) balloon expandable endoprosthesis to the internal iliac arteries.The physician then intubated the patient as there had been a lot of blood loss (amount unknown) from the pseudoanerysm brachial artery bleed in the in the left upper extremity which had caused neuropathy and other issues.When they did the follow up scan prior to doing that? it was discovered that the vbx in left renal artery had migrated out, so they extended that back into the left renal.The physician opted to implant viabahn® first followed by a vbx within the renal as well for more length and to help ¿seat¿ it and help with the tortuosity factor.The patient tolerated the procedure.
|
| |
|
Additional Manufacturer Narrative
|
|
G3: added pma.
|
| |
|
Additional Manufacturer Narrative
|
|
The imaging evaluation performed by a clinical imaging specialist showed the following: ¿ one time-point available for evaluation: post-implantation cta dated (b)(6) 2024.¿ contrast is visualized outside the implanted devices near the celiac artery, sma, and left renal arteries.Contrast is increased on the delayed contrast series imaging.Confirmation of an endoleak of unknown origin in the proximal to mid aneurysm sac.¿ there also appears to be contrast pooling in the anterior aneurysm sac, in the mid to distal sac.The pooling is increased on the delayed contrast series, thereby, suggesting a type ii endoleak possibly involving the ima.All the pooling contrast in the mid-distal sac communicates.¿ there also appears to be contrast in a lumbar artery communicating with the contrast pooling in the sac.Cannot confirm ante grade or retrograde flow with available images.¿ imaging shows stent(s) in the left internal iliac artery and the left external iliac artery.There appears to be seal.No contrast visualized outside the extended stents on the left side.¿ imaging also shows extended stent(s) in the right external iliac artery.There appears to be seal.
|
| |
|