MEDTRONIC IRELAND ONYX FRONTIER; STENT, CORONARY, DRUG-ELUTING Back to Search Results
Model Number ONYXNG35012UX
Medical Device Problem Codes Device Damaged by Another Device (2915); Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/03/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
An attempt was made to use one 3.5x12mm onyx frontier coronary drug eluting stent to treat a mildly tortuous, moderately calcified lesion with 90% stenosis in the proximal-mid-distal left anterior descending (lad) artery.The device was inspected with no issues.Negative prep was not performed.The lesion was pre-dilated.The device did pass through a previously deployed stent.Resistance was not encountered when advancing the device.Excessive force was not used.It was reported that stent dislodgement/displacement occurred during delivery to/at lesion.A non-medtronic guide extension catheter was in use and one 2.75x38mm stent and one 3.0x30mm stent were placed.Multiple balloons were used after stent placement, and it is believed that use of the balloons deformed the proximal edge of the second stent placed.When the 3.5x12mm onyx frontier was being placed, the deformed proximal edge of the second stent caused the onyx frontier to catch when sliding back into place, and the onyx frontier stent started to come off the balloon.The stent was deployed in place at the location and didn't pull completely off the balloon.The patient is alive with no injury.
 
Additional Manufacturer Narrative
Additional information: the 3.5x12mm onyx frontier coronary drug eluting stent was treating a lesion in the proximal left anterior descending (lad) artery.The 2.75x38mm stent and 3.0x30mm stent that were placed were medtronic devices.The proximal edge of the second 3.0x30mm medtronic stent deformed after multiple ballooning.The deformed proximal edge of the 3.0x30mm medtronic stent was treated with ballooning.The patient is alive with no further injury.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Additional Manufacturer Narrative
Additional information: a non-medtronic guide extension catheter was in use and one 2.75x38mm medtronic stent and one 3.0x30mm medtronic onyx frontier stent were placed.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Additional Manufacturer Narrative
Additional information: annex d code.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ONYX FRONTIER
Common Device Name
STENT, CORONARY, DRUG-ELUTING
Manufacturer (Section D)
MEDTRONIC IRELAND
parkmore business park west
galway
EI 
Manufacturer (Section G)
MEDTRONIC IRELAND
parkmore business park west
galway
EI  
Manufacturer Contact
alison sweeney
parkmore business park west
galway 
EI  
091708096
MDR Report Key20133192
Report Number9612164-2024-04034
Device Sequence Number7296445
Product Code NIQ
UDI-Device Identifier00763000511555
UDI-Public00763000511555
Combination Product (Y/N)N
Initial Reporter StateVA
Initial Reporter CountryUS
PMA/510(K) Number
P160043
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2024
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 10/23/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberONYXNG35012UX
Device Catalogue NumberONYXNG35012UX
Device Lot Number0012202217
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/03/2024
Supplement Date Received by Manufacturer09/04/2024
09/18/2024
10/21/2024
Initial Report FDA Received Date09/04/2024
Supplement Report FDA Received Date09/12/2024
10/09/2024
10/23/2024
Date Device Manufactured03/27/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age61 YR
Patient SexMale
Patient Weight81 KG
-
-