• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. ADAPTA; PULSE GENERATOR, PERMANENT, IMPLANTABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. ADAPTA; PULSE GENERATOR, PERMANENT, IMPLANTABLE Back to Search Results
Model Number ADDRS1
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 09/21/2022
Type of Reportable Event Death
Additional Manufacturer Narrative
Concomitant medical products: (b)(6) non-defib lead, implanted on (b)(6) 2008.39565-65 pain stim lead, implanted on (b)(6) 2013.37714 pain stim ipg, implanted on (b)(6) 2013.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
It was reported that the patient died.Follow-up was attempted to obtain the relevant information; however, a cause of death could not be identified.This event is being conservatively reported in an abundance of caution in response to the advisory describing the potential for device circuit error to occur while the device is processing an atrial-sensed episode.This patient was confirmed to be programmed to a pacing mode susceptible to the described advisory issue.
 
Additional Manufacturer Narrative
Product event summary: the device was returned and analyzed.Analysis of the device revealed normal battery depletion.The battery indicator signifying that it is time for device replacement occurred after explant.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADAPTA
Common Device Name
PULSE GENERATOR, PERMANENT, IMPLANTABLE
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer Contact
gerard torenvliet
8200 coral sea st ne
mounds view, MN 55112-4391
7635269277
MDR Report Key20133347
Report Number3004209178-2024-17836
Device Sequence Number7221189
Product Code NVZ
UDI-Device Identifier00643169708747
UDI-Public00643169708747
Combination Product (Y/N)N
Initial Reporter StateMN
Initial Reporter CountryUS
PMA/510(K) Number
P980035
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Remedial Action Recall
Type of Report Initial,Followup
Report Date (Section B) 10/01/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date04/14/2019
Device Model NumberADDRS1
Device Catalogue NumberADDRS1
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/05/2024
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/05/2024
Supplement Date Received by Manufacturer09/12/2024
Initial Report FDA Received Date09/04/2024
Supplement Report FDA Received Date10/01/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/17/2017
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
1688TC LEAD
Outcome Attributed to Adverse Event Death;
Patient Age84 YR
Patient SexFemale
-
-