|
Reported event: an event regarding crack/fracture involving a triathlon insert was reported.The event was not confirmed.Method & results: product evaluation and results: visual inspection of the returned device indicated that the insert is worn.The damage present on the posterior portion was consistent with delamination and material loss.The damage on the articulating surface of the insert is consistent with burnishing, scratching, and third body indentations, which are common damage modes of uhmwpe.The yellow discoloration observed on the tibial insert is consistent with the absorption of synovial fluid by the device.Clinician review: a review of the provided medical records by a clinical consultant indicated: "i can confirm that the patient had a cementless right triathlon tritanium total knee arthroplasty at some point in the past because i was able to see an x-ray with the implant in place, albeit undated.I cannot confirm the revision procedure as i have no office notes, operation notes or post revision x-rays.The root cause of this event cannot be determined with certainty.The causes of polyethylene failure are multifactorial.One factor is surgical technique in the way the implant is positioned and aligned and whether or not proper stability and kinematics of the knee was restored.In this case the tibial component is in significant varus and with a significantly increased posterior slope.These factors alone can contribute to polyethylene failure both at the post and at the locking mechanism.Patient factors including lifestyle, activity level and bmi can also contribute, as well as implant factors" product history review: review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.Complaint history review: there have been no other similar events for the lot referenced.Conclusions: it was reported that the patient was revised due to crack/fracture.Visual inspection of the returned device indicated that the insert is worn.The damage present on the posterior portion was consistent with delamination and material loss.The damage on the articulating surface of the insert is consistent with burnishing, scratching, and third body indentations; which are common damage modes of uhmwpe.The yellow discoloration observed on the tibial insert is consistent with the absorption of synovial fluid by the device.A review of the provided medical records by a clinical consultant indicated: "i can confirm that the patient had a cementless right triathlon tritanium total knee arthroplasty at some point in the past because i was able to see an x-ray with the implant in place, albeit undated.I cannot confirm the revision procedure as i have no office notes, operation notes or post revision x-rays.The root cause of this event cannot be determined with certainty.The causes of polyethylene failure are multifactorial.One factor is surgical technique in the way the implant is positioned and aligned and whether or not proper stability and kinematics of the knee was restored.In this case the tibial component is in significant varus and with a significantly increased posterior slope.These factors alone can contribute to polyethylene failure both at the post and at the locking mechanism.Patient factors including lifestyle, activity level and bmi can also contribute, as well as implant factors"the event was not confirmed.No further investigation for this event is possible at this time.If devices and / or additional information become available to indicate further evaluation is warranted, this record will be reopened.
|