| Catalog Number |
ECS29B |
| Medical Device Problem Code |
Failure to Form Staple (2579)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/26/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during an unknown left colon procedure that about one forth of the staples on the outer rim appeared to be unformed.Leak test was preformed and no leaks were detected.There were no adverse consequences reported.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Date sent 9/4/2024 d4 batch # unk d4: udi: as the lot number for the device involved in the event was not provided, the full udi is currently not available.D4: udi: the expiration date is currently not available.Therefore, the full udi is currently not available.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.No lot or batch number was provided therefore a device history could not be done.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Date sent: 9/17/2024.The device upon which this medwatch is based has been received, however, the evaluation is not yet complete.Any further information derived from the evaluation will be submitted in a supplemental 3500a form.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Date sent: 9/24/2024 d4: batch #770c43 investigation summary the product was returned to ethicon endo-surgery for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample revealed that the ecs29b device arrived with no apparent damage.Only the anvil half washer was present and there were no staples present, indicating that the device achieved a full firing stroke.The device was reloaded with staples, a new washer was placed on the device and it was tested for functionality.It fired and formed all the staples as well as completely cut the test media and the breakaway washer without incident.The staple line was complete, and the staples meet the staple form release criteria.As part of ethicon endo surgery¿s quality process, all devices are manufactured, inspected, and released to approved specifications.A manufacturing record evaluation was performed for the finished device batch number 770c43, and no non-conformances were identified.Although no conclusion could be reached on the cause of the reported event, the instructions for use do contain the following caution: the firing stroke must be completed.Do not partially fire the instrument.Incomplete firing can result in malformed staples, incomplete cut line, bleeding, and leakage from the staple line and/or difficulty removing the device.Ensure that the firing trigger is fully squeezed to ensure proper staple formation and cutting of tissue.The event described could not be confirmed as the device performed without any difficulties.Although no product defect was identified, there may have been other circumstances or issues that occurred during the use of the device that could not be replicated during the laboratory analysis.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post-market surveillance.
|
| |
|