EXACTECH, INC. NV GXL LINER NEUTRAL, 36MM ID, GROUP 2 CUPS; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED
|
Back to Search Results |
|
| Catalog Number |
130-36-52 |
| Medical Device Problem Code |
Naturally Worn (2988)
|
| Health Effect - Clinical Codes |
Emotional Changes (1831); Pain (1994); Loss of Range of Motion (2032); Osteolysis (2377); Swelling/ Edema (4577)
|
| Date of Event |
04/23/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported via legal documentation that approximately 130 months after a left total hip replacement procedure, the patient underwent a revision procedure to address prosthesis wear.Plaintiff, avery knight, suffered from pain, swelling, limtited motion and had various procedures done on his left hip as a remedy to fix the damage inflicted from the implanted exactech device.The implantation of the exactech device and revision has caused plaintiff, avery knight, serious personal injuries, severe pain and suffering, physical disability, mental anguish, emotional distress, fear, loss of enjoyment of life, medical expenses, and financial losses, and he will sustain future damages, including but not limited to cost of medical care, rehabilitation, home health care, loss of earning capacity, mental and emotional distress, and pain and suffering.Accordingly, plaintiff is seeking compensatory and special damages including attorney's fees and costs and all other available remedies under the law.Plaintiff sharon knight has and will continue to suffer loss of consortium, companionship, and loss of services, which will be permanent.No further issues or complications were reported.No additional information is available.
|
| |
|
Additional Manufacturer Narrative
|
|
D10 concomitant devices (b)(6) 170-36-00 - biolox delta femoral head 36mm od, +0mm (b)(6) 180-01-52 - nv crown cup clstr hole 52mm group 2 (b)(6) 164-01-11 - element-stem, collarless w/ha, std offset, sz 11(b)(6) 120-65-25 - bone screw 6.5mm dia x 25mm long (b)(6) 120-65-30 - bone screw 6.5mm dia x 30mm long the product associated with the reported event is within the scope of recall z-1732-2022; however, there is insufficient information to evaluate whether the subject issue of the recall was a cause or contributor to the reported event.The device was not returned for evaluation and no medical or other records containing treatment information or patient information have been received; therefore, the reported event cannot be confirmed, nor can the circumstances or potential causes or contributors to the alleged event be evaluated.Should additional, material information become available that permits more analysis or conclusions, a supplemental report will be filed accordingly.
|
| |
|
Additional Manufacturer Narrative
|
|
Based on the available information, the patient meets the following risk criteria for early prosthesis wear and/or osteolysis: implanted with a lateralized liner, combination of largest available femoral head and/or thinnest available acetabular liner was used and implanted with a component having a shelf age of greater than 2 years.The most likely cause for the reported revision due to prosthesis wear and osteolysis may have been a combination of risk factors such as use error, implant positioning, implant size selection, and patient factors (fitness for surgery, biomechanics, activity level and local tissue oxidation potential).However, this cannot be confirmed as the devices were not returned for evaluation, and images and radiographs were not provided.H6: corrected health effect clinical codes.
|
| |
|