| Catalog Number |
0502904030 |
| Medical Device Problem Codes |
Fogging (1253); Moisture or Humidity Problem (2986)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/17/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Additional information will be provided once the investigation has been completed.The device manufacturer date is not known at this time.However, should it become available it will be provided in future reports.
|
| |
|
Event or Problem Description
|
|
It was reported that there was foggy image.
|
| |
|
Additional Manufacturer Narrative
|
|
This complaint investigation was closed based on the device not received, therefore the reported failure mode was not confirmed.In the event that the device is received, the complaint will be reopened and the investigation will be updated with new results.Alleged failure: fogging.Probable root cause: laser welding seal failure; distal/proximal window solder failure; damage to optical train; damage to needle; damage to distal or proximal windows; moisture intrusion; end of life wear-out; environmental disturbance: endoscope colder than dew point; environmental disturbance: fluid entrapment between the coupler and eyepiece junction (eyepiece only); use error.The reported failure mode will be monitored for future reoccurrence.
|
| |
|
Event or Problem Description
|
|
It was reported that there was foggy image.
|
| |
|
Search Alerts/Recalls
|