U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDLINE INDUSTRIES, LP; CATHETER, CLOSED SUCTION, 8FR, Y

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDLINE INDUSTRIES, LP; CATHETER, CLOSED SUCTION, 8FR, Y Back to Search Results
Catalog Number DYNCPY8
Medical Device Problem Code Leak/Splash (1354)
Health Effect - Clinical Code Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
Date of Event 08/07/2024
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
According to the customer, "the vent wouldn't return volumes while on patient but would work fine on the bag".The customer reported "the inline was preventing the return of volumes", and when the device was changed "everything returned to normal".No additional details are available related to the customer reported issue.Despite multiple good faith efforts to obtain additional information, the customer contact was unable or unwilling to provide further incident details to the manufacturer.No additional information is available.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this medwatch is being filed.If any further relevant information is identified or obtained, a supplemental medwatch will be submitted.
 
Event or Problem Description
According to the customer, "the vent wouldn't return volumes while on patient but would work fine on the bag".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Common Device Name
CATHETER, CLOSED SUCTION, 8FR, Y
Manufacturer (Section D)
MEDLINE INDUSTRIES, LP
three lakes drive
northfield IL 60093 2753
Manufacturer Contact
danielle tognocchi
three lakes dr
northfield, IL 60093
MDR Report Key20136883
Report Number1417592-2024-00838
Device Sequence Number7416937
Product Code BSY
UDI-Device Identifier40884389129907
UDI-Public40884389129907
Combination Product (Y/N)N
Initial Reporter StateAR
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 09/04/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue NumberDYNCPY8
Device Lot Number2862312004
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/07/2024
Initial Report FDA Received Date09/04/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-