• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PHYSIO-CONTROL, INC. - 3015876 LIFEPAK CR2 DEFIBRILLATOR; AUTOMATED EXTERNAL DEFIBRILLATORS (NON-WEARABLE)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PHYSIO-CONTROL, INC. - 3015876 LIFEPAK CR2 DEFIBRILLATOR; AUTOMATED EXTERNAL DEFIBRILLATORS (NON-WEARABLE) Back to Search Results
Model Number CR2
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/05/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The customer received a replacement lid and battery, an engineering investigation has determined that due to the design of the lcpr2 lid switch, absence of the lid magnet allows current to flow from the battery, even when the device is in standby mode.This will reduce the life of the battery.Therefore, the reported issue, the loss of lid/lid magnet, is associated with two hazardous situations: opening the lid does not turn on the device, and, higher than intended current draw while the device is in standby mode results in a prematurely depleted the battery.Replacement of the device lid so that the magnet is present in the device allows the lid switch to function as designed; turning on/off the device when the lid is opened and preventing current draw when the device lid is closed.The lpcr2 operating instructions has been updated to include additional troubleshooting tips to direct the customer to act when device behavior indicates the lid magnet may be missing.A new warning informs the customer of the risk associated with a missing magnet.
 
Event or Problem Description
The customer contacted stryker to report that their device¿s lid magnet was missing from the lid.As a result, the device may not automatically power on when the lid is opened.This may lead to defibrillation therapy being delayed or unavailable, if needed.There was no report of patient use associated with the reported event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIFEPAK CR2 DEFIBRILLATOR
Common Device Name
AUTOMATED EXTERNAL DEFIBRILLATORS (NON-WEARABLE)
Manufacturer (Section D)
PHYSIO-CONTROL, INC. - 3015876
11811 willows road ne
redmond WA 98052
Manufacturer (Section G)
PHYSIO-CONTROL, INC. - 3015876
11811 willows road ne
redmond WA 98052
Manufacturer Contact
brian blakeslee
11811 willows road ne
redmond, WA 98052
4258674000
MDR Report Key20137230
Report Number0003015876-2024-02550
Device Sequence Number7482289
Product Code MKJ
UDI-Device Identifier00883873824108
UDI-Public00883873824108
Combination Product (Y/N)N
Initial Reporter StateAR
Initial Reporter CountryUS
PMA/510(K) Number
P170018
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Non-Healthcare Professional
Remedial Action Recall
Type of Report Initial
Report Date (Section B) 09/04/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberCR2
Device Catalogue Number99512-001264
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 08/05/2024
Initial Report FDA Received Date09/04/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured03/25/2020
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
3015876-01/14/2021-001-C
Patient Sequence Number1
Patient SexUnknown
-
-