| Catalog Number |
03.133.006 |
| Medical Device Problem Codes |
Device-Device Incompatibility (2919); Material Integrity Problem (2978)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/09/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H11: additional narrative: d4: udi: as the lot number for the device involved in the event was not provided, the full udi is currently not available.D9: complainant part is expected to be returned for manufacturer review/investigation, but has yet to be received.E3: initial reporter is a depuysynthes sales representative h6: without a lot number the device history records review could not be completed.Product was not returned.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported during a procedure on (b)(6) 2024, a drill was jammed into the drill guide.They wanted to implant a 2.7 lag screw, so he took on a 2.7 non-locking drill guide placed over a 2.7 drill.He tried to advance/retract the drill in an attempt to dislodge it but was unsuccessful and the sleeve part of the drill guide was also damaged.They opened a new set and used the drill guide from that one.The procedure was successfully completed with a two to three minute delay and no consequence to the patient.This report is for a drill guide.This is report 1 of 2 for (b)(4).
|
| |
|
Event or Problem Description
|
|
Additional information received states that there was no allegation against the lag screw and no other medical intervention was required.
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H6 component codes: most relevant component code is g07002 (appropriate term/code not available) to capture no findings available due to no product returned.H3, h6: product was not returned.Based on the information available, it has been determined that no corrective and preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Search Alerts/Recalls
|