| Model Number |
ZKU150 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Blurred Vision (2137)
|
| Date of Event |
08/07/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Section a3b, a4 and a5: unknown/ asked but not available.Section h3: other 81: the device was not returned for evaluation.Therefore, a failure analysis of the complaint device cannot be completed.A review of the device history record, complaint trending, and risk documentation for this device will be performed.Upon completion of the review, if there is any further relevant information a supplemental medwatch will be filed.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that two preloaded intraocular lenses(iol) were implanted and explanted from the left eye of the patient.The first lens (zku150u185, s/n (b)(6)) was implanted in the patient's eye and explanted due to blurry vision and replaced with same model and diopter lens (s/n:(b)(6)) in a secondary procedure.The patient complained of blurry vision with the replaced second lens too.The replaced second lens was explanted and replaced with lens from another manufacturer in a secondary procedure.There were no patient issues.Platinum cartridge was used.No other information is available.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: section d9: device available for evaluation: yes.Section d9: returned to manufacturer on: 08/12/2024.Section h3: device evaluated by manufacturer: yes.Device evaluation: visual inspection of the complaint lens reveal that it was received cut into three pieces with one piece missing.The lens pieces were cleaned and no further issues were identified.Conclusion: the complaint issues "blurry vision" and "explant" were not identified during product evaluation.The other observed issues during the product evaluation could not be confirmed to be related to a manufacturing or design issue.Therefore, there is no indication of a product deficiency or product malfunction.Based on the complaint investigation results, the product was released within specifications.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
|
| |
|
Search Alerts/Recalls
|