• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PHYSIO-CONTROL, INC. - 3015876 LIFEPAK® 15 DEFIBRILLATOR/MONITOR; AUTOMATED EXTERNAL DEFIBRILLATORS (NON-WEARABLE)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PHYSIO-CONTROL, INC. - 3015876 LIFEPAK® 15 DEFIBRILLATOR/MONITOR; AUTOMATED EXTERNAL DEFIBRILLATORS (NON-WEARABLE) Back to Search Results
Model Number 15
Medical Device Problem Code Unexpected Shutdown (4019)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/08/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device was returned to stryker's product analysis center (pac) for evaluation.The pac was able to verify the reported issue through review of software logs but unable to duplicate during testing.The battery pins were proactively replaced.After completing other unrelated repairs, proper device operation was observed through functional and performance testing.The device was returned to the customer for use.The cause of the reported issue could not be determined.Stryker contacted the customer to request additional information on the patient.Stryker requests patient information necessary for regulatory compliance, strictly adhering to hipaa's privacy rule.No response has been received from the customer.Patient fields in which information is not provided were intentionally left blank.
 
Event or Problem Description
The customer contacted stryker to report their device unexpectedly lost power.N this state the device may not be able to deliver defibrillation therapy if needed.This issue is patient related; however there was no adverse event reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIFEPAK® 15 DEFIBRILLATOR/MONITOR
Common Device Name
AUTOMATED EXTERNAL DEFIBRILLATORS (NON-WEARABLE)
Manufacturer (Section D)
PHYSIO-CONTROL, INC. - 3015876
11811 willows road ne
redmond WA 98052
Manufacturer (Section G)
PHYSIO-CONTROL, INC. - 3015876
11811 willows road ne
redmond WA 98052
Manufacturer Contact
brian blakeslee
11811 willows road ne
redmond, WA 98052
4258674000
MDR Report Key20143117
Report Number0003015876-2024-02559
Device Sequence Number15303424
Product Code MKJ
UDI-Device Identifier00883873871713
UDI-Public00883873871713
Combination Product (Y/N)N
Initial Reporter StateSD
Initial Reporter CountryUS
PMA/510(K) Number
P160026
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 09/05/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number15
Device Catalogue Number99577-001955
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/13/2024
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 08/08/2024
Initial Report FDA Received Date09/05/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/16/2022
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-