| Catalog Number |
138110 |
| Medical Device Problem Code |
Delivered as Unsterile Product (1421)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Received three 138110 in original packaging.Lot number was verified.Performed a visual inspection, there were no obvious signs of a breach in the seal; however, the device was encroaching into the seal.Performed a functional inspection, the devices were dye leak tested which indicated that the packaging had an insufficient heat seal on two of the packages.A likely cause for this issue may be due to shipping and handling causing the device to encroach into the seal and/ or the seal at the point of encroachment was weak allowing the device to penetrate the seal.The manufacturing documents from the device history record have been reviewed with special attention to the manufacturing and inspection of the product.The product released for distribution was found to have met all specifications prior to shipment.A 2 year lot history review shows a total of 6 devices for this lot number and failure mode.A two-year review of complaint history revealed there has been a total of (b)(4) reports, regarding (b)(4) devices, for this device family and failure mode.During this same time frame (b)(4) devices have been manufactured and shipped worldwide.Should all the complaint devices have been found confirmed for this reported failure, the rate of failure would be (b)(4).Per the instructions for use, the user is advised the following: inspect and test each device before each use.We will continue to monitor for trends through the complaint system to assure patient safety.
|
| |
|
Event or Problem Description
|
|
During incoming inspection, the distributor rejected this device, 138110, elec,blde,lng,ext,insulatd,40, for an insufficient heatseal.There was no contact with the patient as this was found during incoming inspection.This report is being raised due to the reported malfunction with potential for injury upon reoccurrence.
|
| |
|
Search Alerts/Recalls
|
|
|