AESCULAP AG AS VEGA PS FEMORAL COMP.CEMENTED F6R; KNEE ENDOPROSTHETICS Back to Search Results
Model Number NX034Z
Medical Device Problem Codes Patient Device Interaction Problem (4001); Loosening of Implant Not Related to Bone-Ingrowth (4002)
Health Effect - Clinical Code Failure of Implant (1924)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Investigation on-going.Should relevant additional information / investigation results become available, a supplemental medwatch report will be submitted.
 
Event or Problem Description
It was reported that there was an issue with a component of the vega knee implant system.According to the complaint description, the revision surgery was planned for (b)(6) 2024 for an unspecified reason.The polyethylene components would be removed and replaced.The implantation on the right knee had occurred in 2022.Additional information was not provided but has been requested.The adverse event is filed under aesculap aesculap reference number (b)(4).
 
Additional Manufacturer Narrative
Investigation on-going.Should relevant additional information / investigation results become available, a supplemental medwatch report will be submitted.
 
Event or Problem Description
Additional information reported that the reason for the surgery was due to a screw coming lose in the construct, as such, the implant was floating around in the joint.The adverse event is filed under aesculap reference number (b)(4).2916714-2024-00130 (aic reference no.(b)(4)).2916714-2024-00148 (aic reference no.(b)(4)).2916714-2024-00149 (aic reference no.(b)(4)).2916714-2024-00150 (aic reference no.(b)(4)).
 
Additional Manufacturer Narrative
Investigation results: the device was not returned to the manufacturer for physical evaluation.Additionally, no on-site evaluation of the unit was performed and no parts were returned for failure analysis.Therefore, the investigation was not able to confirm a device issue that could be associated with the reported event.Batch history review: the device history records (dhr) were not able to be reviewed as no lot number was made available.However, all device history records (dhr) are reviewed and released according to documented procedures and a device is not released if it does not meet requirements or is nonconforming.Conclusion and measures / preventative measures: a definitive conclusion regarding the complaint incident cannot be reached without a physical examination of the complaint device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AS VEGA PS FEMORAL COMP.CEMENTED F6R
Common Device Name
KNEE ENDOPROSTHETICS
Manufacturer (Section D)
AESCULAP AG
po box 40
tuttlingen, 78501
GM  78501
MDR Report Key20143995
Report Number2916714-2024-00130
Device Sequence Number20908214
Product Code JWH
UDI-Device Identifier04046963810221
UDI-Public04046963810221
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2024
Serviced by Third Party (Y/N)Unknown
Reporter Type Importer
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 11/18/2024
Report Date (Section F) 09/05/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberNX034Z
Device Catalogue NumberNX034Z
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Distributor Facility Aware Date08/12/2024
Event Location Hospital
Date Report to Manufacturer08/12/2024
Initial Date Received by Manufacturer Not provided
Supplement Date Received by ManufacturerNot provided
Not provided
Initial Report FDA Received Date09/05/2024
Supplement Report FDA Received Date10/21/2024
11/18/2024
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexMale
Date Report Sent to FDA09/05/2024
-
-