• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SYNTHES GMBH SYNFRAME RING CLAMP; ORTHOPEDIC MANUAL SURGICAL INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNTHES GMBH SYNFRAME RING CLAMP; ORTHOPEDIC MANUAL SURGICAL INSTRUMENT Back to Search Results
Catalog Number 387.347
Medical Device Problem Code Compatibility Problem (2960)
Health Effect - Clinical Code Insufficient Information (4580)
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that on an unknown date the team is having an issue with the synframe clamps where the clamps are losing their grip strength on the frame.It is unknown when the issue was discovered.It is unknown if there were patient and procedure involvement.This report is for one synframe ring clamp for (b)(4).
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H11 additional narrative: b3: unknown event date d4: udi: as the lot number for the device involved in the event was not provided, the full udi is currently not available.H3, h4, h6: without a lot number the device history records review could not be completed.Product was not returned.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H11 additional narrative: added: d4 (lot), d9, h4 corrected: d4 (primary udi number) if information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H11 additional narrative: h3, h4, h6 the product was returned to j&j medtech orthopaedics for evaluation.Visual inspection of the returned sample found the device exhibits signs repeated usage like scratches and small dents over the attaching ring surface.A functional test was performed with a sample mating device (product code: 387.337) and the device was affixed properly.However, the interaction was found slightly loose.Consequently, the reported condition can be confirmed.The observed condition was identified as an end of life indicator; damage consistent with repeated use and servicing.The lifecycle requirements of the device are event related and depend on the use and inspection of the device in clinical practice.As the device can be damaged on the first or 100th use, the device must be properly inspected prior to each surgical use.Refer to the device/country specific ifu for information related to end of life, reprocessing instructions, and inspection procedures.A dimensional inspection was not performed since it was not applicable to the complaint condition.The overall complaint was confirmed as the observed condition of the synframe ring clamp would have contributed to the complained issue.Based on the investigation findings, the potential cause is traced to end of life, and it has been determined that no corrective and/or preventative action is required.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of j&j medtech orthopaedics quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history review: part# 387.347, lot # l790788, manufacturing site:, werk hägendorf, supplier ; na, release to warehouse date: 29 jun 2018, expiration date: na.A manufacturing record evaluation was performed for the finished article lot and no non-conformances were identified if information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNFRAME RING CLAMP
Common Device Name
ORTHOPEDIC MANUAL SURGICAL INSTRUMENT
Manufacturer (Section D)
SYNTHES GMBH
luzernstrasse 19-21
zuchwil 4528
SZ  4528
Manufacturer (Section G)
WERK HÄGENDORF (CH)
im bifang 6
haegendorf 4614
SZ   4614
Manufacturer Contact
kate karberg
1302 wright lane east
west chester, PA 19380
3035526892
MDR Report Key20144028
Report Number8030965-2024-10747
Device Sequence Number8396852
Product Code LXH
UDI-Device Identifier10705034771137
UDI-Public(01)10705034771137(10)L790788
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Distributor
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number387.347
Device Lot NumberL790788
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 08/09/2024
Supplement Date Received by Manufacturer12/23/2024
01/09/2025
Initial Report FDA Received Date09/05/2024
Supplement Report FDA Received Date12/27/2024
01/13/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/29/2018
Is the Device Labeled for Single Use? (Y/N) No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-