Model Number 6887900
Medical Device Problem Code Output Problem (3005)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/02/2024
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the anesthesia system failed the pressure transducer test during system checkout.There was no patient involvement.Manufacturer's reference #: (b)(4).
 
Additional Manufacturer Narrative
On-site investigation was performed by our field service engineer.The claimed issue was reproduced during troubleshooting.According to received service report, replacement of the inspiratory pressure transducer printed circuit board solved the issue.The ventilator passed all functional and safety tests and was cleared for clinical use.The pressure transducer printed circuit board is part of the pressure measuring in the system.The device's logs were received and evaluation of the test log shows that pressure transducer test failed several times on provided date of event.Our conclusion is that the replaced part was the cause of the failure.However, the root cause to why the part failed has not been established as the replaced part was not returned for investigation.
 
Event or Problem Description
Manufacturer's reference #: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FLOW-E ANESTHESIA SYSTEM
Common Device Name
GAS-MACHINE, ANESTHESIA
Manufacturer (Section D)
MAQUET CRITICAL CARE AB
roentgenvagen 2
solna
Manufacturer (Section G)
MAQUET CRITICAL CARE AB
roentgenvagen 2
solna
Manufacturer Contact
caroline kabbabe
roentgenvagen 2
solna 
MDR Report Key20144588
Report Number8010042-2024-01452
Device Sequence Number11831018
Product Code BSZ
UDI-Device Identifier07325710010457
UDI-Public(01)07325710010457(11)210205
Combination Product (Y/N)N
Initial Reporter CountryIT
PMA/510(K) Number
K191027
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 09/05/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number6887900
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/02/2024
Supplement Date Received by Manufacturer10/03/2024
Initial Report FDA Received Date09/05/2024
Supplement Report FDA Received Date10/04/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/05/2021
Is the Device Labeled for Single Use? (Y/N) No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-