| Model Number |
6887900 |
| Medical Device Problem Code |
Output Problem (3005)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/02/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the anesthesia system failed the pressure transducer test during system checkout.There was no patient involvement.Manufacturer's reference #: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
On-site investigation was performed by our field service engineer.The claimed issue was reproduced during troubleshooting.According to received service report, replacement of the inspiratory pressure transducer printed circuit board solved the issue.The ventilator passed all functional and safety tests and was cleared for clinical use.The pressure transducer printed circuit board is part of the pressure measuring in the system.The device's logs were received and evaluation of the test log shows that pressure transducer test failed several times on provided date of event.Our conclusion is that the replaced part was the cause of the failure.However, the root cause to why the part failed has not been established as the replaced part was not returned for investigation.
|
| |
|
Event or Problem Description
|
|
Manufacturer's reference #: (b)(4).
|
| |
|
Search Alerts/Recalls
|