| Catalog Number |
ECS25B |
| Medical Device Problem Code |
Failure to Fire (2610)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
08/26/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).D4 batch # unk.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.The manufacturing records were reviewed and the manufacturing/packaging criteria were met prior to the release of this lot.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported that during a robotic esophagectomy procedure that when the fellow went to fire the first stapler she didn't hear the "crunch" when the device was opened it was found the device cut but did not staple.Procedure was converted to open.The staple line was then over sewed.Patient currently doing ok.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Date sent: 10/3/2024.D4: batch #908c35.Investigation summary: the product was returned for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample determined that the ecs25b device arrived with no apparent damage, there were no staples present in the device, the breakaway washer uncut and without marks.The device was reloaded with staples and tested with the returned washer for functionality; the device formed all the staples, as well as completely cut the test media and the breakaway washer without incident.The staple line was complete and the staples meet the staple form release criteria.The condition of the washer indicates that the device had not been fired through a full firing stroke, the orange indicator was not fully into the safe green firing range or the anvil was not reattached correctly.It should be noted that before firing, ensure that the orange indicator is fully within the green range of the gap setting scale and the anvil is reattached correctly.To reattach the detachable head or anvil do not clamp across or grip on the locking springs as it might not click into place.Not reattaching the anvil correctly or the orange indicator not being set on the green range will result in a bigger gap between the anvil and the guide face, therefore the staples will not hit the anvil to make b- formed staples.In addition, if the firing sequence is not complete, staples could be partially deployed without cutting the washer.As part of ethicon endo surgery¿s quality process, all devices are manufactured, inspected, and released to approved specifications.A manufacturing record evaluation was performed for the finished device batch number 908c35, and no non-conformances were identified.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Date sent: 10/24/2024.Additional information received: both the fellow and dr.Have extensive experience with our manual circular staplers and do not feel it was user error.He has since upgraded to powered circular.
|
| |
|