| Model Number |
N/A |
| Medical Device Problem Code |
Loosening of Implant Not Related to Bone-Ingrowth (4002)
|
| Health Effect - Clinical Codes |
Failure of Implant (1924); Pain (1994); Scar Tissue (2060); Ambulation or Postural Difficulties (2544)
|
| Date of Event |
10/30/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that the patient underwent a left knee arthroplasty revision of the femoral and articular surface components to address pain, instability, scar tissue and femoral implant loosening approximately three (3) years following the patient's last knee arthroplasty.Initial operative notes noted no intraoperative complications.Revision operative notes note the patient was undergoing a left knee revision with extensive synovectomy and release of scar tissue as well as application of long leg splint under anesthesia.The femoral components and polyethylene were exchanged without complications.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).D10 - concomitant devices - nexgen cemented option femoral component left size f catalog #: 00599401691 lot #: 63376531, nexgen straight stem extension 17mm x 100mm catalog #: 00598801017 lot #: 62319367n, nexgen posterior femoral augment block precoat size f 5mm catalog #: 00599003601 lot #: 63088982, nexgen distal femoral augment block precoat size f 10mm catalog #: 00599003620 lot #: 62917296, nexgen distal femoral augment block precoat size f 10mm catalog #: 00599003620 lot #: 62917296, nexgen articular surface with locking screw green 10mm size e, f catalog #: 00599404010 lot #: 62227678, nexgen stemmed precoat tibial component size 6 catalog #: 00598004702 lot #: 63377478, nexgen tibial block and screws size 6 5mm catalog #: 00598800626 lot #: 63101473, nexgen straight stem extension 12.7mm x 30mm catalog #: 00598801212 lot #: 62719567, nexgen tibial block and screws size 6 5mm catalog #: 00598800626 lot #: 63108231, nexgen standard all poly patella size 35mm catalog #: 00597206535 lot #: 61113009.The complainant has indicated that the product will not be returned to zimmer biomet for investigation.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
|
| |
|
Event or Problem Description
|
|
No further event information available at the time of this report.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay additional information.The following sections were updated: b4, g3, g6, h2, h3, h6, h11.Medical records provided confirmed the patient's reported instability and pain.Radiographic review of x-rays provided identified anatomic alignment and progressive osteopenia with no signs of loosening, wear, radiolucency or abnormalities.The device history records were reviewed and no discrepancies were identified.A definitive root cause could not be determined with the information provided.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|