AIZU OLYMPUS CO., LTD. OES BRONCHOFIBERSCOPE Back to Search Results
Model Number BF-P60
Medical Device Problem Codes Peeled/Delaminated (1454); Device Markings/Labelling Problem (2911)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Event or Problem Description
It was observed during the device inspection, that the bronchofiberscope had tube coating peeled, and insertion section flexible tube display disappeared.There were no reports of patient involvement.
 
Additional Manufacturer Narrative
The device evaluation is ongoing.Should additional relevant information become available, a supplemental report will be submitted.
 
Additional Manufacturer Narrative
This supplemental report is being submitted to provide the results of the legal manufacturer's final investigation.A review of device history record and historical complaints analysis was conducted and no deviations that could have caused or contributed to the reported issue was identified.The device was returned to olympus for inspection.A definitive root cause could not be determined.Based on the results of the investigation, it is likely that the following led to the malfunction: due to stress from use, external factors, or handling.Olympus will continue to monitor the field performance of this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OES BRONCHOFIBERSCOPE
Common Device Name
BRONCHOFIBERSCOPE
Manufacturer (Section D)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA  965-8520
Manufacturer (Section G)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key20146715
Report Number9610595-2024-17760
Device Sequence Number8259039
Product Code EOQ
UDI-Device Identifier04953170339196
UDI-Public04953170339196
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K023984
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 09/18/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model NumberBF-P60
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/22/2024
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/27/2024
Supplement Date Received by Manufacturer09/11/2024
Initial Report FDA Received Date09/05/2024
Supplement Report FDA Received Date09/18/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured04/18/2011
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-