| Model Number |
WB991046 |
| Medical Device Problem Code |
Output Problem (3005)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/14/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device evaluation is ongoing.Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported, the electrosurgical system generator had an e231 (electrosurgical generator malfunction) error code.The issue occurred, during maintenance.There were no reports of patient harm.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation and device evaluation.Correction: d9.Additional information: h3, h6, h11.The device was returned to olympus for inspection and the reportable malfunction was reproduced.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, the reported error was caused by the printed circuit board/communication board.Olympus will continue to monitor field performance for this device.
|
| |
|
Search Alerts/Recalls
|