• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COVIDIEN MANSFIELD TRUCLEAR; HYSTEROSCOPE (AND ACCESSORIES)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COVIDIEN MANSFIELD TRUCLEAR; HYSTEROSCOPE (AND ACCESSORIES) Back to Search Results
Model Number 7209808
Medical Device Problem Code Mechanical Jam (2983)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 02/08/2022
Type of Reportable Event Malfunction
Event or Problem Description
According to the reporter, during procedure, the device will not go into oscillate mode.There was no patient injury.
 
Additional Manufacturer Narrative
This report is being submitted as part of remediation activities associated with capa # 643143, which is addressing mdr reporting gu idance from fdas 1995 preamble, which ensures complaints related to corrections and recalls which were previously reported to fda under 21 cfr 806 are now reported as mdrs; this is not a new malfunction/event.H3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.The analysis noted that the unit had a digital pcb failure.The issue was resolved by replacing the digital pcb.It was reported that the device would not oscillate.The reported issue was confirmed.The most likely cause was traced to device design.Internal process improvements have been initiated to mitigate this issue.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRUCLEAR
Common Device Name
HYSTEROSCOPE (AND ACCESSORIES)
Manufacturer (Section D)
COVIDIEN MANSFIELD
15 hampshire street
mansfield MA 02048
Manufacturer (Section G)
COVIDIEN MANSFIELD
15 hampshire street
mansfield MA 02048
Manufacturer Contact
justin ellis
8200 coral sea st ne
mounds view, MN 55112
7635265677
MDR Report Key20147379
Report Number1282497-2024-00362
Device Sequence Number4236733
Product Code HIH
UDI-Device Identifier10884521749931
UDI-Public10884521749931
Combination Product (Y/N)N
Initial Reporter StateMN
Initial Reporter CountryUS
PMA/510(K) Number
K031787
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Remedial Action Recall
Type of Report Initial
Report Date (Section B) 09/05/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number7209808
Device Catalogue Number7209808
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/21/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/06/2022
Initial Report FDA Received Date09/05/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
Z-1467-2022
Patient Sequence Number1
Patient SexUnknown
-
-