|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.New information was added to the following fields: h3, h4, and h6.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, olympus confirmed the first event that foreign material was inside of the device, but the type of the material cannot be identified.Based on the investigation result below, the user possibly could not remove the foreign material due to physical damage of the device.However, a definitive root cause cannot be determined.Based on the results of the investigation, olympus confirmed the second event that there was a gap at the distal end and it is likely that the event occurred due to physical or chemical stress applied to the distal end.However, a definitive root cause cannot be determined.The 1st event can be prevented by following the instructions for use which state: -reprocessing manual_ cleaning, disinfection and sterilization procedures precleaning and manual cleaning.The 2nd event can be prevented by following the instructions for use which state: -operation manual_ important information ¿ please read before use_ warnings and cautions: warning: do not strike, hit, or drop the endoscope¿s distal end, insertion tube, bending section, control section, universal cord, or endoscope connector.Also, do not bend, pull, or twist the endoscope¿s distal end, insertion tube, bending section, control section, universal cord, or endoscope connector with excessive force.The endoscope may be damaged and could cause patient injury, burns, bleeding, and/or perforations.It could also cause parts of the endoscope to fall off inside the patient the 1st and 2nd event can be prevented by following the instructions for use which state: -operation manual_ preparation and inspection of the endoscope_ inspection of the endoscope inspect the objective lens at the distal end of the endoscope¿s insertion tube for scratching, cracks, stains, gaps around the lens or other irregularities.-reprocessing manual_ compatible reprocessing methods and chemical agents_ compatibility summary: the endoscope and accessories are compatible with several methods of reprocessing.However, not all reprocessing methods are compatible with all endoscopes and all accessories.Reprocessing with incompatible methods can cause equipment damage even if the number of reprocessing cycles is small.Three attempts were performed to obtain additional information regarding reprocessing steps, but no response was received from the customer.Olympus will continue to monitor field performance for this device.
|