MEDIVANCE, INC. ¿ 1725056 ARCTIC SUN® TEMPERATURE MANAGEMENT SYSTEM, FLUID DELIVERY LINES; ARCTIC SUN FLUID DELIVERY LINE
|
Back to Search Results |
|
| Catalog Number |
734-07 |
| Medical Device Problem Code |
Gas/Air Leak (2946)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/28/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.Udi format updated with available product information per gudid.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Event or Problem Description
|
|
It was reported that they were trying to initiate therapy using arctic sun device but were getting alert 02 (low flow).Flow rate (fr) was 0lpm, inlet pressure (ip) was -0.4psi, circulation pump command (cpc) was 100 percentage, pump hours were 5528, system hours were 6456.Had nurse stop therapy, empty pads, and disconnect pads.Placed device in manual mode, with only the fluid delivery line (fdl) attached, flow rate (fr) was 1.8lpm, inlet pressure (ip) was -7.3psi, circulation pump command (cpc) was 74 percentage.Connected left thigh pad, flow rate (fr) was 2.1lpm, inlet pressure (ip) was -4.6psi.Moved left thigh pad to another valve set, flow rate (fr) was 0.8lpm, inlet pressure (ip) was -7.1psi, circulation pump command (cpc) was 45 percentage.Connected left chest pad, nurse mistakenly attached it to the first valve set location they connected the left thigh pad to at first.0.1lpm, -0.7psi.Moved to the other valve set on that side of the fluid delivery line (fdl), flow rate (fr) was 1.8lpm, inlet pressure (ip) was -7.3psi, circulation pump command (cpc) was 68 percentage.Connected right thigh pad, 2.4lpm, -7.3psi, circulation pump command (cpc) was 84 percentage.Connected right chest pad, flow rate (fr) was 2.9lpm, inlet pressure (ip) was -7.2psi, 96 percentage.Placed device in automatic mode and disabled manual mode, flow rate (fr) was 3lpm.Asked nurse to mark the valve set that seems to be damaged so nurses did not use that.Send device to biomed after use for inspection and repair, if needed.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was not returned.The reported issue was inconclusive.Although a root cause could not be definitively identified, based on the risk documentation review, a potential root cause for this type of failure could be fittings between external components become damage due to repeated force exposure.However, there was insufficient information to confirm this potential root cause.The latest labeling revision was reviewed and found to be adequate.The ifu currently instructs the user on the proper method to use this device to avoid undue injury to the patient and damage to the product.A dhr review is not required as the serial number is unknown.Udi format updated with available product information per gudid.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Event or Problem Description
|
|
It was reported that they were trying to initiate therapy using arctic sun device but were getting alert 02 (low flow).Flow rate (fr) was 0lpm, inlet pressure (ip) was -0.4psi, circulation pump command (cpc) was 100 percentage, pump hours were 5528, system hours were 6456.Had nurse stop therapy, empty pads, and disconnect pads.Placed device in manual mode, with only the fluid delivery line (fdl) attached, flow rate (fr) was 1.8lpm, inlet pressure (ip) was -7.3psi, circulation pump command (cpc) was 74 percentage.Connected left thigh pad, flow rate (fr) was 2.1lpm, inlet pressure (ip) was -4.6psi.Moved left thigh pad to another valve set, flow rate (fr) was 0.8lpm, inlet pressure (ip) was -7.1psi, circulation pump command (cpc) was 45 percentage.Connected left chest pad, nurse mistakenly attached it to the first valve set location they connected the left thigh pad to at first.0.1lpm, -0.7psi.Moved to the other valve set on that side of the fluid delivery line (fdl), flow rate (fr) was 1.8lpm, inlet pressure (ip) was -7.3psi, circulation pump command (cpc) was 68 percentage.Connected right thigh pad, 2.4lpm, -7.3psi, circulation pump command (cpc) was 84 percentage.Connected right chest pad, flow rate (fr) was 2.9lpm, inlet pressure (ip) was -7.2psi, 96 percentage.Placed device in automatic mode and disabled manual mode, flow rate (fr) was 3lpm.Asked nurse to mark the valve set that seems to be damaged so nurses did not use that.Send device to biomed after use for inspection and repair, if needed.
|
| |
|