Model Number VTICM5_12.1
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Blurred Vision (2137)
Date of Event 04/10/2023
Type of Reportable Event Serious Injury
Event or Problem Description
A facility representative reported that following an implantable collamer lens (icl) implantation procedure, the patient experienced refractive surprise (vision discomfort and poor near vision).In a separate visit the lens was explanted and later replaced with a different power lens.The problem resolved.Cause of the event was reported as patient related factor, however it was reported that the device failed to perform as intended.
 
Additional Manufacturer Narrative
H6 - type of investigation - lens work order search: no similar complaint type events reported for units within the same lot.H11- manufacturer narrative: each lens is subjected to a 100% assessment of the power (spherical and cylinder) and optical resolution during the manufacturing process in order to determine acceptability per the lens model and diopter.Secondary surgical intervention to remove/replace/reposition the lens is identified in the labeling as a known potential adverse event following icl implantation.Claim #: (b)(4).
 
Additional Manufacturer Narrative
H6- device evaluation: lens was returned in liquid, in vial.Visual inspection found no visible damage to the lens.Dimensional and functional inspection found the lens to be within specifications.Claim # (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IMPLANTABLE COLLAMER LENS (ICL)
Common Device Name
PHAKIC TORIC INTRAOCULAR LENS
Manufacturer (Section D)
STAAR SURGICAL COMPANY
1911 walker avenue
monrovia CA 91016
Manufacturer (Section G)
STAAR SUGICAL COMPANY
1911 walker avenue
monrovia CA 91016
Manufacturer Contact
joselene muniz
1911 walker avenue
monrovia, CA 91016
6263037902
MDR Report Key20148621
Report Number2023826-2024-04037
Device Sequence Number11831053
Product Code QCB
UDI-Device Identifier00840311304400
UDI-Public00840311304400
Combination Product (Y/N)N
Initial Reporter CountrySN
PMA/510(K) Number
P030016
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Device Explanted Year2023
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/12/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date08/31/2025
Device Model NumberVTICM5_12.1
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/26/2024
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/12/2024
Supplement Date Received by Manufacturer10/15/2024
Initial Report FDA Received Date09/05/2024
Supplement Report FDA Received Date11/11/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/19/2022
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age48 YR
Patient SexMale
-
-