| Model Number |
VTICM5_12.1 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Blurred Vision (2137)
|
| Date of Event |
04/10/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
A facility representative reported that following an implantable collamer lens (icl) implantation procedure, the patient experienced refractive surprise (vision discomfort and poor near vision).In a separate visit the lens was explanted and later replaced with a different power lens.The problem resolved.Cause of the event was reported as patient related factor, however it was reported that the device failed to perform as intended.
|
| |
|
Additional Manufacturer Narrative
|
|
H6 - type of investigation - lens work order search: no similar complaint type events reported for units within the same lot.H11- manufacturer narrative: each lens is subjected to a 100% assessment of the power (spherical and cylinder) and optical resolution during the manufacturing process in order to determine acceptability per the lens model and diopter.Secondary surgical intervention to remove/replace/reposition the lens is identified in the labeling as a known potential adverse event following icl implantation.Claim #: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
H6- device evaluation: lens was returned in liquid, in vial.Visual inspection found no visible damage to the lens.Dimensional and functional inspection found the lens to be within specifications.Claim # (b)(4).
|
| |
|