(b)(4).Date sent: 9/5/2024.D4: batch #362c44.Investigation summary: the product was returned for evaluation.Visual inspection was conducted on the returned reload.Visual analysis of the returned sample determined that the cr40g reload was received with all staples protruding, the washer uncut and with the knife recessed below the reload deck.The drivers were noted to be partially advance.No functional test was performed due to the condition of the reload.The condition of the reload is consistent with the device being partially activated.As a result of the partial firing the lockout was engaged when the device was reopened.Do not fire the instrument unless the closure trigger is properly latched against the handle.The firing trigger must be pulled back completely against the closure trigger to properly fire the instrument.In addition, if the firing sequence is not complete, you could deploy the staples without cutting the washer and forming the staples.Please reference instructions for use for additional information.It should be noted that as part of our quality process al devices are manufactured, inspected, and released to approved specifications.A manufacturing record evaluation was performed for the finished device batch number 362c44, and no non-conformances were identified.The lot#236c16 history records were reviewed and certified by external manufacturing that the manufacturing criteria were met prior to the release of the equipment.The certificate records are accessible through external manufacturing.Attempts are being made to obtain the following information.To date no response has been provided.If further details are received at a later date a supplemental medwatch will be sent: 1.Could you please clarify if there were any patient consequences? 2.Could you please clarify if there was any change in the post-operative care of the patient as a result of the event?.
|