Catalog Number SMHB775
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Capsular Contracture (1761); Hematoma (1884)
Date of Event 03/07/2024
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
On september 05, 2024, the mentor analysis lab received the suspect medical device for evaluation.On september 13, 2024, mentor completed an evaluation on the returned device.Mentor conducted a visual inspection of the device.During visual evaluation, no apparent damage or visual anomalies were observed on the breast implant device.Mentor concluded that the capsular contracture in the patient´s breast was the result of the body´s individual physiological response to the implantation of a foreign object in soft tissue.Capsular contracture is a known complication associated with these devices and is referenced in our current product insert data sheet.The american society of plastic surgeons recommends and encourages member surgeons to always submit breast implants, capsules, and effusion to pathology for examination.A hematoma is a collection of blood within the space around the implant.Symptoms from hematoma may include swelling, pain, and bruising.If a hematoma occurs, it will most likely occur soon after surgery; however, it can occur at any time, such as after an injury to the breast.Small hematomas may be absorbed by the body, while others may require surgery for drainage.Hematoma is a known complication associated with this procedure and is referenced in our product insert data sheet.Manufacturer¿s reference number: (b)(4).
 
Additional Manufacturer Narrative
Since the device has not been returned for analysis, no product failure analysis can be conducted, and no determination of possible contributing factors can be made.  as such, the investigation will be closed.  if the complaint device is received in the future, the investigation will be reopened and conducted as appropriate.A manufacturing record evaluation (mre) was performed, and no anomalies were found related to this complaint.In addition, the mre verifies that the device was manufactured in accordance with documented specification and procedures.Mentor is submitting this report pursuant to the provisions of 21 cfr, part 803.  this report may be based on information which mentor has not been able to investigate or verify prior to the required reporting date.  this report does not reflect a conclusion by fda, mentor, or its employees that the report constitutes an admission that the device, mentor, or its employees caused or contributed to the potential event described in this report.  reason for device explant and/or reoperation: capsular contracture.Manufacturer¿s reference number: (b)(4).
 
Event or Problem Description
It was reported that a patient underwent a breast augmentation revision surgery with implantation of a 775cc mentor memorygel boost breast implant prosthesis.Post-operatively, the patient developed a left breast hematoma.Resultantly, she developed left breast asymmetry and a high sitting left breast.During a physical exam with a medical professional, she was diagnosed with left breast baker grade iii capsular contracture.As a result, the patient underwent bilateral removal and replacement with 730cc mentor memorygel boost breast implants on (b)(6) 2024.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MENTOR MEMORYGEL BOOST BREAST IMPLANT
Common Device Name
PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED
Manufacturer (Section D)
MENTOR TEXAS
3041 skyway circle north
irving TX 75038
Manufacturer (Section G)
MENTOR TEXAS
3041 skyway circle north
irving TX 75038
Manufacturer Contact
kate karberg
3041 skyway circle north
irving, TX 75038
3035526892
MDR Report Key20150918
Report Number1645337-2024-10321
Device Sequence Number4248573
Product Code FTR
UDI-Device Identifier10081317031311
UDI-Public(01)10081317031311(17)280613(10)9912062
Combination Product (Y/N)N
Initial Reporter StateNV
Initial Reporter CountryUS
PMA/510(K) Number
P030053
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Device Explanted Year2024
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue NumberSMHB775
Device Lot Number9912062
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 08/12/2024
Supplement Date Received by Manufacturer09/05/2024
Initial Report FDA Received Date09/05/2024
Supplement Report FDA Received Date09/20/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/17/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age63 YR
Patient SexFemale
-
-