• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BECTON, DICKINSON & CO., (BD) BD VACUTAINER® SAFETY-LOK¿ BLOOD COLLECTION SET PRE-ATTACHED HOLDER; BLOOD SPECIMEN COLLECTION DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON, DICKINSON & CO., (BD) BD VACUTAINER® SAFETY-LOK¿ BLOOD COLLECTION SET PRE-ATTACHED HOLDER; BLOOD SPECIMEN COLLECTION DEVICE Back to Search Results
Catalog Number 368652
Medical Device Problem Code Product Quality Problem (1506)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/06/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A device evaluation and/or device history review is anticipated but is not complete.Upon completion, a supplemental report will be filed.
 
Event or Problem Description
It was reported that while using bd vacutainer® safety-lok¿ blood collection set pre-attached holder, the needle didn't go into the patient's skin smoothly and the sheath starts to retract to cover the needle while still in the patient's vein on (1) device.No patient or user impact reported.
 
Additional Manufacturer Narrative
The following fields were updated due to additional information: d9: device available for evaluation: yes h.3 device eval by manufacturer? yes d9: returned to manufacturer on: 19-aug-2024 the following fields were updated due to additional information: d.4 udi#: (b)((4).Investigation summary bd received one (1) sample for investigation.The sample was evaluated by visual examination for all cannula defects, including needle point integrity, and the indicated failure mode for needle point dull with the incident lot was not observed as all product specifications were met.Additionally, the 1 customer sample was subjected to a test for lube coverage, the sample passed testing exhibiting sufficient lube coverage on the iv cannula and no indication of excessive lube relating to difficult to operate.Based on a review of the device history record for the incident lot, all product specifications and requirements for lot release were met.There were no related quality issues during manufacturing of the product.This complaint is unable to be confirmed for the indicated failure mode of needle point dull and difficult to operate based on the customer returned sample.Bd was not able to identify a root cause for the indicated failure mode.Complaints received for this device and reported condition will continue to be tracked and trended.Our business team regularly reviews the collected data for identification of emerging trends.
 
Event or Problem Description
It was reported that while using bd vacutainer® safety-lok¿ blood collection set pre-attached holder, the needle didn't go into the patient's skin smoothly and the sheath starts to retract to cover the needle while still in the patient's vein on (1) device.No patient or user impact reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD VACUTAINER® SAFETY-LOK¿ BLOOD COLLECTION SET PRE-ATTACHED HOLDER
Common Device Name
BLOOD SPECIMEN COLLECTION DEVICE
Manufacturer (Section D)
BECTON, DICKINSON & CO., (BD)
1575 airport road
sumter SC 29153
Manufacturer (Section G)
BECTON, DICKINSON & CO., (BD)
1575 airport road
sumter SC 29153
Manufacturer Contact
jo doyka
7 loveton circle
sparks, MD 21152
4103164000
MDR Report Key20152333
Report Number1024879-2024-00787
Device Sequence Number8397594
Product Code JKA
UDI-Device Identifier30382903686521
UDI-Public(01)30382903686521(11)240318(17)260131(10)4037624
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
K980414
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 11/12/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date01/31/2026
Device Catalogue Number368652
Device Lot Number4037624
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/06/2024
Supplement Date Received by Manufacturer11/12/2024
Initial Report FDA Received Date09/05/2024
Supplement Report FDA Received Date11/22/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/18/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-