| Catalog Number |
ECH45L |
| Medical Device Problem Codes |
Delivered as Unsterile Product (1421); Appropriate Device Problem Term/Code Not Available (3191)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/10/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during the lap appendectomy procedure that upon opening the device a "moisture" like film was inside the clear plastic sleeve that covers most of the jaw.Two devices were opened in the procedure with this "moisture" like film.A third like device was used to continue with the procedure.There were no adverse consequences reported.After the procedure their inventory was checked and another three were found to have this substance on them as well.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).D4: udi: the expiration date is currently not available.Therefore, the full udi is currently not available.An analysis of the product could not be performed since a physical sample was not received for evaluation.An evaluation of the manufacturing record could not be performed as the required product identification number was not provided to complete the evaluation.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Date sent: 9/24/2024 d4: batch # c9d99x investigation summary the product was returned to ethicon endo surgery for evaluation.Visual inspection were conducted on the returned device.Visual analysis of the returned sample revealed that one ech45l device was returned inside its package unopened; upon visual inspection, excess of lubricant was noted in the jaws and the joint cover area.In addition, no issues were observed related to integrity and the seal area was noted completed.During the manufacturing process, a clear lubricant is applied to the articulation joint cover to make the insertion and extraction of the device into the trocar easier.The lubricant is sodium stearate; this is known to be matter non-toxic and biologically non-hazardous that is attributable to the assembly process.This lubricant is safe for patient use.Based on the information currently available, a product issue was identified during the investigation of the sample received.This product issue will be addressed through ethicon endo surgery¿s quality system.Device history review a manufacturing record evaluation was performed for the finished device lot number c9d99x, and no non-conformances were identified.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Date sent: 10/14/2024.Corrected data: h6 medical device problem code.Upon review of the information provided, it was concluded that this event does not meet the fda defined criteria for a malfunction, and has been revised to a not reportable event.
|
| |
|