• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BECTON, DICKINSON & CO., (BD) BD VACUTAINER® ECLIPSE¿ BLOOD COLLECTION NEEDLE WITH PRE-ATTACHED HOLDER; HYPODERMIC SINGLE LUMEN NEEDLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON, DICKINSON & CO., (BD) BD VACUTAINER® ECLIPSE¿ BLOOD COLLECTION NEEDLE WITH PRE-ATTACHED HOLDER; HYPODERMIC SINGLE LUMEN NEEDLE Back to Search Results
Catalog Number 368651
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/08/2024
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the safety shield is detaching from an unspecified number of bd vacutainer® eclipse¿ blood collection needles with pre-attached holder.This is occurring both prior to and during use.There was no report of adverse impact to patients or users.
 
Additional Manufacturer Narrative
Investigation summary: "material #: 368651.Lot/batch #: 4065938.Bd had not received samples, but 1 photo was provided for investigation.The photo was reviewed and the indicated failure mode for defective locking mechanism was observed.Additionally, 20 retention samples from bd inventory were evaluated by functional testing, each having their safety shield engaged, and the issue of defective locking mechanism was not observed.Based on a review of the device history record for the incident lot, all product specifications and requirements for lot release were met.There were no related quality issues during manufacturing of the product.This complaint has been confirmed for the indicated failure mode defective locking mechanism.Bd was not able to identify a root cause for the indicated failure mode.Complaints received for this device and reported condition will continue to be tracked and trended.Our business team regularly reviews the collected data for identification of emerging trends.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD VACUTAINER® ECLIPSE¿ BLOOD COLLECTION NEEDLE WITH PRE-ATTACHED HOLDER
Common Device Name
HYPODERMIC SINGLE LUMEN NEEDLE
Manufacturer (Section D)
BECTON, DICKINSON & CO., (BD)
1575 airport road
sumter SC 29153
Manufacturer (Section G)
BECTON, DICKINSON & CO., (BD)
1575 airport road
sumter SC 29153
Manufacturer Contact
jo doyka
7 loveton circle
sparks, MD 21152
4103164000
MDR Report Key20153981
Report Number1024879-2024-00790
Device Sequence Number19895542
Product Code FMI
UDI-Device Identifier50382903686518
UDI-Public(01)50382903686518
Combination Product (Y/N)N
Initial Reporter CountrySW
PMA/510(K) Number
K982541
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 08/13/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Catalogue Number368651
Device Lot Number4065938
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/08/2024
Initial Report FDA Received Date09/05/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured03/05/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-