Received two 138105 in original packaging.Lot number was verified.Performed a visual inspection, there were no obvious signs of a breach in the seal however, the device was encroaching into the seal.Performed a functional inspection, the devices were dye leak tested which indicated that the packaging had an insufficient heat seal on one of the packages.Root cause cannot be determined, however, based upon evaluation of the device, a possible cause of this event is the device was encroaching into the seal.The manufacturing documents from the device history record have been reviewed with special attention to the manufacturing and inspection of the product.The product released for distribution was found to have met all specifications prior to shipment.A two-year lot history review shows a total of two (2) devices for this lot number and failure mode.A two-year review of complaint history revealed there has been a total of 223 complaints, regarding 1,196 devices, for this device family and failure mode.During this same time frame 8,964,158 devices have been manufactured and shipped worldwide.Should all the complaint devices have been found confirmed for this reported failure, the rate of failure would be 0.0001.Per the instructions for use, the user is advised to inspect and test each device before each use.We will continue to monitor for trends through the complaint system to assure patient safety.
|
|
During incoming inspection, the distributor rejected this device, 138105, for an insufficient heatseal.There was no contact with the patient as this was found during incoming inspection.This will be reported as a malfunction with potential for injury upon reoccurrence.
|