| Model Number |
N/A |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
Foreign Body In Patient (2687)
|
| Date of Event |
07/17/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The manufacturer has received the sample and will be evaluated.A supplemental will be submitted with evaluation results.
|
| |
|
Event or Problem Description
|
|
It was reported that the complete break of the catheter with retention of 39cm of catheter.The retained piece of line was retrieved via cut down in ir.Line had a securacath securement device.Line failure point was between the 6 and 7 cm markers on the line.No other information was provided.
|
| |
|
Event or Problem Description
|
|
It was reported that the complete break of the catheter with retention of 39cm of catheter.The retained piece of line was retrieved via cut down in ir.Line had a securacath securement device.Line failure point was between the 6 and 7 cm markers on the line.No other information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The complaint of a break in the catheter was confirmed.The product returned for evaluation was one 4 fr s/l powerpicc solo catheter.The investigation findings are consistent with damage accumulated through flexural fatigue.Flexural fatigue occurs due to cyclic kinking of the catheter tube in which physiological, placement, usage, and mechanical factors may gradually form a crack(s) in the catheter.The returned product sample was evaluated and a break was observed at the 7 cm depth marker.The catheter break contained physical features associated with material fatigue, and the characteristics observed which supported this type of failure included: ¿ damage which was circumferentially aligned.¿ fracture edges which were rounded and polished due to repeated material wear.¿ 'c' shaped impressions leading into the fracture site which are consistent with material buckling due to movement which caused the fracture edges to be pressed together.¿ overall elliptical shape to the fracture cross-section (a result of repeated kinking of the tubing).An examination of the catheter structure revealed no potential damage/defect related to manufacture of the product.The damage location suggested that catheter securement, access and maintenance techniques may have contributed.This complaint will be recorded for future trending and monitoring purposes.
|
| |
|