|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, and although we could not specify the root cause of the suggested event, it is likely the event was caused by stress applied on the insertion section.The event can be detected by following the instructions for use which state: 3.2 inspection of the endoscope 5.Inspect the external surface of the entire insertion tube including the bending section and the distal end for dents, bulges, swelling, peeling, scratching, holes, sagging, transformation, bends, adhesion of foreign body, dropout of parts, any protruding objects or other irregularities.Olympus will continue to monitor field performance for this device.
|