|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, it is likely that the error and probe breakage occurred due to contact with a surgical instrument or the non-insulated area of the grasping section.The possible mechanisms are as follows: during seal & cut mode output activation, the probe came into contact with hard tissue, metal, or a surgical instrument, resulting in a scratch on the probe and causing the seal & cut short circuit error.Seal & cut mode was activated when the grasping section was closed without grasping tissue (this includes after tissue resection).This resulted in wear on the tissue pad.Due to the wear of the tissue pad, the non-insulated area of the grasping section came into contact with the probe.In addition, a seal & cut short circuit error (error code: u508) occurred.Seal & cut mode was activated under the described condition, and scratches (contact marks) were generated when the probe contacted the distal end of the grasping section.Olympus will continue to monitor field performance for this device.
|