| Catalog Number |
4722 |
| Medical Device Problem Code |
Material Fragmentation (1261)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
08/09/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that the air + broke during surgery.Broken anchor pieces were potentially unable to be retrieved.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information will be provided once the investigation has been completed.The device manufacturer date is not known at this time.However, should it become available it will be provided in future reports.
|
| |
|
Event or Problem Description
|
|
It was reported that the air + broke during surgery.Broken anchor pieces were potentially unable to be retrieved.
|
| |
|
Additional Manufacturer Narrative
|
|
This complaint investigation was closed based on the device not received, therefore the reported failure mode was not confirmed.In the event that the device is received, the complaint will be reopened and the investigation will be updated with new results.Alleged failure: anchor broken.Probable root cause: application: - user not familiar with device use.- user excessively bends/kinks the needle.The reported failure mode will be monitored for future reoccurrence.Manufacture date is not known.
|
| |
|