| Model Number |
102620-81A |
| Medical Device Problem Code |
Failure to Sense (1559)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/29/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The manufacturer is currently performing an investigation and will provide the results with the supplemental report.
|
| |
|
Event or Problem Description
|
|
On august 7, 2024, senseonics was made aware of an incident where the user experienced sensor inaccuracy which led to early sensor removal.
|
| |
|
Additional Manufacturer Narrative
|
|
The sensor replacement alert was first presented to the user on (b)(6) 2024, day 136 after insertion.The sensor was tested in-house, and the review of investigation analysis revealed a loss of chemical performance.The system correctly disabled the sensor when it detected the performance failure, and the system's self-test functions were working normally.The root cause of the performance failure was due to the sensor's hydrogel oxidation.As part of resolution, a return material authorization was issued for sensor replacement.No further resolution was necessary for this complaint.B4.Date of this report (b)(6) 2024.D9 device available for evaluation? yes, device received on 09/24/2024.G3 date received by the manufacturer? 08 october 2024.H3.Device evaluated by manufacturer? yes.H6.Type of investigation updated to 10.H6.Investigation findings updated to 3231.H6.Investigation conclusions updated to 4307.
|
| |
|