Investigation conclusion: retained devices from the reported lot number were tested with hcg-negative clinical urine samples.The results were read at 3 and 4 minutes and all devices yielded the expected negative results.No false positive results were observed during in-house testing.The case details were reviewed along with the complaint history for the reported issue and no indications of a systemic issue were identified.Manufacturing batch record review did not uncover any relevant non-conformance's and found that the lot met quality control specifications.Review of the risk management report for this product found that the reported issue is within the risk profile for this device; no new hazard has been identified.A root cause could not be determined from the available information as the reported issue was not replicated during testing of retained product.However, complaint details indicate that urinalysis of the patient samples showed elevated levels of blood, leukocytes, urobilinogen, protein, ketones, and ascorbic acid.While hemoglobin and ascorbic acid did not interfere with the assay when tested at concentrations of 1,000 mg/dl and 20 mg/dl, respectively, the other substances have not been evaluated for interference with this assay.Please see the interfering substances section of the package insert for additional information.Complaints are tracked and trended on a monthly basis.Per the package insert: very low levels of hcg (less than 50 miu/ml) are present in urine specimen shortly after implantation.However, because a significant number of first trimester pregnancies terminate for natural reasons, a test result that is weakly positive should be confirmed by retesting with a first morning urine specimen collected 48 hours later.A number of conditions other than pregnancy, including trophoblastic disease and certain non-trophoblastic neoplasms including testicular tumors, prostate cancer, breast cancer, and lung cancer, cause elevated levels of hcg.Therefore, the presence of hcg in urine should not be used to diagnose pregnancy unless these conditions have been ruled out.This test provides a presumptive diagnosis for pregnancy.A confirmed pregnancy diagnosis should only be made by a physician after all clinical and laboratory findings have been evaluated.
|
|
The customer reported a false positive result when testing a patient urine sample using the one step+ hcg urine strip test.The patient presented with abdominal pain and hematuria.The patient was sent to the emergency department for additional work-up, where it was determined that the patient was not pregnant.It was later stated that the patient had been hospitalized, however no additional details were provided.Attempts to obtain further information are ongoing.The customer reported a false positive result for an additional patient.See h10 for the related mdr number.
|