| Model Number |
ES-3870K |
| Medical Device Problem Code |
Erratic or Intermittent Display (1182)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/19/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The product was returned to pentax medical for repair.Our technician checked the returned unit and confirmed that the ccd module with drive pcb distorted image.Based on the result, we concluded that it was caused due to the excessive force applied on the ccd module with drive pcb.Based on the technical report ""hr-rpt-0586(image failure)"" and/or the risk analysis results, it was evaluated to submit mdr.
|
| |
|
Event or Problem Description
|
|
The time of event is unknown.There was no report of patient harm.Video image failure.
|
| |
|
Search Alerts/Recalls
|