MDT POWERED SURGICAL SOLUTIONS LEGEND METAL CUTTER ATTACHMENT; MOTOR, DRILL, PNEUMATIC Back to Search Results
Model Number ASMC
Medical Device Problem Codes Difficult to Insert (1316); Physical Resistance/Sticking (4012)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/20/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
H3: product analysis:evaluation determined that flat compression spring is missing.The likely cause of failure is inadequate preventive maintenance.It was also noted that not possible to lock a bur inside the attachment because the locking mechanism does not work because of a damaged drive shaft, laser markings are starting to fade, the colorband is discoloredand the bearings feel rough, they are worn.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
Repair request initiated for device with the report of dissecting tool does not attach.It was reported that there was no patient involvement.Repair escalated to product event on evaluation due to detached component.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEGEND METAL CUTTER ATTACHMENT
Common Device Name
MOTOR, DRILL, PNEUMATIC
Manufacturer (Section D)
MDT POWERED SURGICAL SOLUTIONS
4620 n beach st
fort worth TX 76137
Manufacturer (Section G)
MDT POWERED SURGICAL SOLUTIONS
4620 n beach st
fort worth TX 76137
Manufacturer Contact
justin ellis
8200 coral sea street ne mvs21
mounds view, MN 55112
7635265677
MDR Report Key20167774
Report Number1625507-2024-00583
Device Sequence Number20583597
Product Code HBB
UDI-Device Identifier00613994418814
UDI-Public00613994418814
Combination Product (Y/N)N
Initial Reporter CountryEI
PMA/510(K) Number
K020069
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 09/09/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberASMC
Device Catalogue NumberASMC
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/15/2024
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/12/2024
Initial Report FDA Received Date09/09/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/20/2016
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-