H3: product analysis:evaluation determined that flat compression spring is missing.The likely cause of failure is inadequate preventive maintenance.It was also noted that not possible to lock a bur inside the attachment because the locking mechanism does not work because of a damaged drive shaft, laser markings are starting to fade, the colorband is discoloredand the bearings feel rough, they are worn.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|