| Model Number |
RETROCONSTRUCTION CONSTANT FEMORAL GUIDE |
| Medical Device Problem Code |
Physical Resistance/Sticking (4012)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/19/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Investigation is in process.A follow-up report will be provided upon availability of additional information.
|
| |
|
Event or Problem Description
|
|
It was reported that during a shoulder prothesis surgery the 3mm drill was passed through the drill sleeve and the devices then cold-welded.This resulted in abrasion.There was no harm for patient, operator or third party.The surgery was finished successfully with a different device.It was not necessary to switch the surgical technique or do a second surgery.***update (b)(6) 30-aug-2024: the abrasion did not enter the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
Complaint allegation is confirmed.One unpackaged ar-1778r-30 serial/batch number (b)(6) was received for evaluation.Functional testing was not performed because the device arrived with other components stuck inside.Visual inspection revealed scratches and lines around the shaft outside diameter, an indication of abrasions, and the device being in contact with another instrument during use.The most likely cause of the reported failure is misuse due to user error.
|
| |
|