• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: EXACTECH, INC. NV GXL LINR, NTRL, 32MM ID, GROUP 2 CUPS; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EXACTECH, INC. NV GXL LINR, NTRL, 32MM ID, GROUP 2 CUPS; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED Back to Search Results
Catalog Number 130-32-52
Medical Device Problem Codes Unstable (1667); Naturally Worn (2988)
Health Effect - Clinical Codes Pain (1994); Osteolysis (2377); Joint Laxity (4526)
Date of Event 09/05/2012
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
D10: concomitant devices: 2427287 180-01-52 - nv crown cup clstr hole 52mm group 2.2330886 164-01-11 - element-stem, collarless w/ha, std offset, sz 11.2044785 142-32-00 - cocr fem head 32mm +0 offset 12/14.2388330 122-65-25 - 6.5mm acetabular bone screw 25mm.The product associated with the reported event is within the scope of recall z-1729-2022; however, there is insufficient information to evaluate whether the subject issue of the recall was a cause or contributor to the reported event.The device was not returned for evaluation and no medical or other records containing treatment information or patient information have been received; therefore, the reported event cannot be confirmed, nor can the circumstances or potential causes or contributors to the alleged event be evaluated.Should additional, material information become available that permits more analysis or conclusions, a supplemental report will be filed accordingly.
 
Event or Problem Description
It was reported that approximately 123 months after a right total hip replacement procedure, the patient underwent a revision procedure to address loosening, severe pain; severe osteolysis, significant polyethylene wear, cup had minimal fixation to bone.Revision surgery operative notes were provided.Patient left the operating room in stable condition.Post operative diagnosis noted wear of the implant.No further issues or complications were reported.No additional information is available.
 
Additional Manufacturer Narrative
Multiple mdr reports were filed for this event, please see associated reports: 1038671-2024-03365.This follow-up report is being submitted to relay additional and/or corrected information.The most likely underlying cause for the revision reported is a combination of risk factors specified such as use error, implant positioning, implant size selection, and patient factors (fitness for surgery, biomechanics, activity level and local tissue oxidation potential) and being implanted for more than 10 years.However, this cannot be confirmed from the reported information and the devices were not available for evaluation.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NV GXL LINR, NTRL, 32MM ID, GROUP 2 CUPS
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED
Manufacturer (Section D)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer (Section G)
EXACTECH INC.
2320 nw 66th court
gainesville 32653
Manufacturer Contact
geoff gannon
2320 nw 66th court
gainesville 32653
3523771140
MDR Report Key20169404
Report Number1038671-2024-03363
Device Sequence Number8069690
Product Code JDI
UDI-Device Identifier10885862022165
UDI-Public10885862022165
Combination Product (Y/N)N
Initial Reporter StateAR
Initial Reporter CountryUS
PMA/510(K) Number
K070479
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2012
Device Explanted Year2022
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Consumer,Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/12/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/19/2017
Device Catalogue Number130-32-52
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/03/2024
Supplement Date Received by Manufacturer02/11/2025
Initial Report FDA Received Date09/09/2024
Supplement Report FDA Received Date02/12/2025
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured07/23/2012
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H11
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient SexFemale
-
-