| Model Number |
977C165 |
| Medical Device Problem Code |
High impedance (1291)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/20/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Information was received from a manufacturer representative regarding a patient who was implanted with an implantable neurostimulator (ins).It was reported that during the implant procedure the lead was found to have multiple contacts over 40000 ohms.The lead was tried in the implant and two different wireless external neurostimulators and all had the same result.There were no other stimulation issues.The cause was not known and the issue was resolved.
|
| |
|
Additional Manufacturer Narrative
|
|
Continuation of d10: product id neu_siliconeanchor (serial: unknown); product type: 0001-accessory; implant date ; explant date product id neu_siliconeanchor (serial: unknown); product type: 0001-accessory; implant date ; explant date medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Additional Manufacturer Narrative
|
|
H3: analysis of the electrical stimulation lead (serial number (b)(6)) found no anomaly.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Search Alerts/Recalls
|