• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SYNTHES GMBH 4.5 VA-LCP CRVD COND PL/12HOLE/266/LT-S; PLATE, FIXATION, BONE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNTHES GMBH 4.5 VA-LCP CRVD COND PL/12HOLE/266/LT-S; PLATE, FIXATION, BONE Back to Search Results
Catalog Number 02.124.413S
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Nonunion/Delayed-union Bone Fracture (2369)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H11 additional narrative: d4: udi: as the lot number for the device involved in the event was not provided, the full udi is currently not available.H3, h4, h6: without a lot number the device history records review could not be completed.Product was not returned.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Subject id: (b)(6).It was reported clinical adverse events were received for breakage of screws and non-union.Event occurred between 3 months and 12 months postop.Patient status/ outcome: patient recovered without persistent damage this report is for one 4.5 va-lcp crvd cond pl/12hole/266/lt-s for (b)(6).(b)(4) is related to (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
4.5 VA-LCP CRVD COND PL/12HOLE/266/LT-S
Common Device Name
PLATE, FIXATION, BONE
Manufacturer (Section D)
SYNTHES GMBH
luzernstrasse 19-21
zuchwil 4528
SZ  4528
Manufacturer (Section G)
MONUMENT
1101 synthes avenue
monument CO 80132
Manufacturer Contact
kate karberg
1302 wright lane east
west chester, PA 19380
3035526892
MDR Report Key20173708
Report Number8030965-2024-11019
Device Sequence Number20867474
Product Code HRS
UDI-Device Identifier10886982042880
UDI-Public(01)10886982042880
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K110354
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Study,Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue Number02.124.413S
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 08/13/2024
Initial Report FDA Received Date09/09/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Unknown
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-