The product was returned to pentax medical for repair.Our technician checked the returned unit and confirmed that the suction nipple stuck accessory/object.Based on the result, we concluded that it was caused due to the inadequate/insufficient reprocessing at the facility on the suction nipple.In addition, our technician confirmed that the insertion flexible tube fluid damage, the segment fluid damage, the operation channel (primary) leak, the distal body worn out, the lcb (light carrying bundle) defective, the suction channel kink, and the insertion flexible tube dented; however, these defects are not the main cause, and/or irrelevant to the alleged complaint.Based on the technical report ""hr-rpt-0588(channel)"" and/or the risk analysis results, it was evaluated to submit mdr.
|